Click for PDF
Technical Data Sheet
Eastman Tritan™ copolyester DA001


Applications
  • Dental aligners and arches
  • Medical devices
Key Attributes
  • Chemical and stain resistant
  • Does not contain Bisphenol-A (BPA)
  • Does not contain plasticizers
  • Excellent clarity
  • Suitable for gamma and E-beam irradiation and ethylene oxide sterilization

Product Description
Eastman Tritan™ DA001 is an amorphous copolyester that offers excellent clarity, toughness, flexibility and chemical resistance needed for use in medically extruded film applications. Film and sheet manufactured from Tritan™ DA001 can be thermoformed with a wide processing window that allows for product designs that reflect intricate detail.

This product also meets ISO 10993 and/or USP Class VI biocompatibility requirement.


Typical Properties

PropertyaTest MethodbTypical Value, Unitsc
Film Properties
Thickness of Film Tested 760 Microns (30 mils)
Tensile Stress @ Yield 
     M.D.D 63835 MPa (5100 psi)
Elongation @ Yield 
     M.D.D 6385 %
Tensile Stress @ Break 
     M.D.D 63844 MPa (6400 psi)
Elongation @ Break 
     M.D.D 638150 %
Youngs Modulus 
     M.D.D 6381000 MPa (1.45 x 105 psi)
Flexural Yield StressD 79044 MPa (6400 psi)
Flexural ModulusD 7901700 MPa (2.47 x 105 psi)
Tear Resistance, Trouser @ 200 mm/minD 193834 N/mm
Mechanical Properties (Injection Molded), ASTM Method
Specific Gravity 
     @ 22°CD 7921.19 g/cm3
Water Absorption, 24 h immersionD 5704.9 %
Izod Impact Strength, Notched @ 23°C (73°F)D 2561180 J/m
Izod Impact Strength, Unnotched @ 23°C (73°F)D 2562030 J/m
Rockwell Hardness, R ScaleD 78596
Thermal Properties
Glass Transition Temperature (Tg)DSC86 °C (187 °F)
Deflection Temperature 
     @ 0.455 MPa (66 psi)D 64875 °C (167 °F)
     @ 1.82 MPa (264 psi)D 64864 °C (147 °F)
Typical Drying Conditions
Drying Temperature 66 °C (150 °F)
Moisture Content Upon Drying <300 ppm
Typical Processing Conditions
Processing Melt Temperature 250-260 °C (480-500 °F)

aUnless noted otherwise, all tests are run at 23°C (73°F) and 50% relative humidity.
bUnless noted otherwise, the test method is ASTM.
cUnits are in SI or US customary units.


Comments
Properties reported here are typical of average lots. Eastman makes no representation that the material in any particular shipment will conform exactly to the values given.


Eastman and its marketing affiliates shall not be responsible for the use of this information, or of any product, method, or apparatus mentioned, and you must make your own determination of its suitability and completeness for your own use, for the protection of the environment, and for the health and safety of your employees and purchasers of your products. No warranty is made of the merchantability of fitness of any product, and nothing herein waives any of the Seller's conditions of sale.


6/2/2025 10:14:18 AM